Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. ANGIOGRAPHY CATH LAB, Medline Kit Number/SKUDYNJ23532C; 2. CATH LAB PACK, Medline Kit Number/SKUDYNJ65185; 3. ANGIO KIT, Medline Kit Number/SKUVASC1057. Recall

Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. ANGIOGRAPHY CATH LAB, Medline Kit Number/SKUDYNJ23532C; 2. CATH LAB PACK, Medline Kit Number/SKUDYNJ65185; 3. ANGIO KIT, Medline Kit Number/SKUVASC1057.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. ANGIOGRAPHY CATH LAB, Medline Kit Number/SKUDYNJ23532C; 2. CATH LAB PACK, Medline Kit Number/SKUDYNJ65185; 3. ANGIO KIT, Medline Kit Number/SKUVASC1057.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2143-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Mar 24, 2026
FDA report date
May 20, 2026
Quantity in commerce
508 kits
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. ANGIOGRAPHY CATH LAB, Medline Kit Number/SKUDYNJ23532C; 2. CATH LAB PACK, Medline Kit Number/SKUDYNJ65185; 3. ANGIO KIT, Medline Kit Number/SKUVASC1057.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Medline Kit Number/SKU DYNJ23532C: UDI/DI each 10193489460872, UDI/DI case 40193489460873, Lot Number: 25DMA809; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 23GBA252; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 23HBI873; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 23IBR043; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 23KBF277; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 24ABQ959; Medline Kit Number/SKU VASC1057: UDI/DI each 10889942431922, UDI/DI case 40889942431923, Lot Number: 25KMF194.

View the original FDA recall notice

Why was this product recalled?

Medline has identified the presence of particulate within the fluid path of the Manifolds.

What actions did Medline Industries, LP take?

Medline Industries, LP issued an URGENT MEDICAL DEVICE notice to its consignees on 3/24/2026 via email and mail. The notice explained the issue, potential risk and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-046 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.