Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma. Recall

Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2152-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 12, 2025
FDA report date
Jul 30, 2025
Quantity in commerce
496 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.

Affected model & UDI checker

Search the exact identifiers published for Z-2152-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
DSL8700Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.Check source noticeNot specified
Not specifiedEstrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.Check source notice15099590211615
FDA code information

UDI: 15099590211615/Lot: 250106D, 250203C, 250203D, 250303C, 250331C, 250428C, 250526C

View the original FDA recall notice

Why was this product recalled?

A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.

What actions did BECKMAN COULTER, INC. take?

On June 12, 2025, Beckman Coulter issued a "Urgent: Medical Device Recall" notification via Email. On July 8, 2025, Beckman Coulter issued a updated notification to provide additional information on the health risk and actions consignees should take. Beckman Coulter ask consignees to take the following actions: 1. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 2. Please discard any remaining inventory of effective lots per your laboratory protocols and ask for replacement. 2. Please share this information with your laboratory staff and retain this notifications as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. 3. Please respond within 10 days in one of the following ways: 1. Electronically, if you received the notification via email, 2. Manually, complete and return the enclosed Response Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.