Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2162-2026

Diowave Laser System, REF: Diowave 250W Recall

Diowave Laser System, REF: Diowave 250W

FDA recall Class II

Ongoing

FDA lists this Class II recall for Diowave Laser System, REF: Diowave 250W. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2162-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 9, 2026
FDA report date
May 20, 2026
Quantity in commerce
16
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Diowave Laser System, REF: Diowave 250W

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI-DI: G180DIOWAVESTEALTHMAX0. All software versions prior to, but not including V1.81

View the original FDA recall notice

Why was this product recalled?

Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthritis, and to temporarily increase local blood circulation, operating at a maximum average power output of 70W, which is beyond the FDA cleared 60W.

What actions did Technological Medical Advancements LLC take?

Starting on 1/9/2026, customer notifications were sent informing customers to install software update V1.81. On 3/11/2026, correction notices were sent to customers informing them to take the following actions: 1) Install Software Update V1.81. Follow the WiFi upgrade instructions to complete the update. The update is free of charge and requires no hardware modifications. 2) Confirm Completion. Once the update is installed, please complete and return the enclosed Correction Response Form by email to info@diowavelaser.com. Alternatively, mail the completed form to TMA at the address above. 3) Forward This Notice. This notice must be passed on to all persons within your organization who need to be aware of this correction, and to any organization where the affected device has been transferred or is currently in use. 4) Contact Support If Needed. Firm's technical support team is available Monday through Friday, 9:00 AM 5:00 PM ET, at (561) 722-1153 or info@diowavelaser.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.