Microbore Extension Set with 0.2 Micron Filter. Recall

Microbore Extension Set with 0.2 Micron Filter. Catalog Number: V6215. for use with IV tubing.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Microbore Extension Set with 0.2 Micron Filter.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2176-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jul 9, 2025
FDA report date
Aug 13, 2025
Quantity in commerce
34,825 units US
Distribution
See key details
FDA cause
Process control

What is being recalled?

Microbore Extension Set with 0.2 Micron Filter. Catalog Number: V6215. for use with IV tubing.

Affected model & UDI checker

Search the exact identifiers published for Z-2176-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
V6215Microbore Extension Set with 0.2 Micron Filter.Check source noticeNot specified
Not specifiedMicrobore Extension Set with 0.2 Micron Filter.Check source notice04046964189173
Not specifiedMicrobore Extension Set with 0.2 Micron Filter.Check source notice04046964189166
FDA code information

Catalog Number: V6215. Primary UDI-DI 04046964189173, Unit of Use UDI-DI: 04046964189166. Lot Number (Expiration Date): 0061742452 (30JUN2025), 0061747379 (31AUG2025 0, 0061767411 (31JAN2026), 0061780914 (30APR2026), 0061803499 (30SEP2026), 0061806173 (30NOV2026), 0061822333 (30APR2027), 0061836578 (30APR2027), 0061849109 (31JUL2027), 0061850836 (31AUG2027), 0061899902 (30SEP2028), 0061936368 (31MAR2029), 0061936119 (31MAR2029), 0061936369 (31MAR2029), 0061940584 (30APR2029).

View the original FDA recall notice

Why was this product recalled?

Label is incorrect. The product label indicates that the device contains an air eliminating filter. However, the filter utilized on this extension set is not indicated for removal of air and does not feature an air vent, which is common to air eliminating filters.

What actions did B. BRAUN MEDICAL INC. take?

B. Braun Medical notified consignees of the recall on about 07/09/2025 via "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter dated 07/02/2025. The letter described the product, problem and actions to be taken. Consignees were instructed to ensure all users within your organization as well as customers, if further distributed, are notified of the recall and provided the notification letter. Review inventory and determine if any affected units are on hand, cease use and quarantine product-Do not destroy any affected product. Complete and return the Urgent Medical Device Recall Acknowledgement Form via fax to (610) 849-1197 or email to: recalls@bbraunusa.com and arrange for the return of affected units. Once the firm receives the Acknowledgement form a customer support representative will contact you with instructions on how to return the product. For questions about this recall, please contact our BBMI's Recalls Department at 844-903-6417.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology