Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2178-2025

Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems Recall

Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems

FDA recall Class II

Ongoing

FDA lists this Class II recall for Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2178-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 23, 2025
FDA report date
Aug 6, 2025
Quantity in commerce
59 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI: 00724995171155: Serial numbers: 0416920048, 0417820084, 0417820088, 0417820096, 0417820110, 0417820092, 0417820100, 0418420047, 0419620004, 0419520013, 0428620109, 0428620112, 0430020191, 0431720044, 0435120043, 0435120047, 0435120051, 0401321042, 0413221047, 0413221199, 0413721008, 0413721012, 0417221023, 0417221027, 0417221031, 0420021018, 0420021022, 0420021026, 0420721021, 0429720003, 0420921073, 0426321186, 0426321190, 0427121121, 0427121125, 0427321012, 0427321002, 0427321005, 0413921017, 0434121095, 0434121098, 0434121101, 0434121104, 0434121107, 0434121110, 0434121113, 0400722081, 0406022021, 0406022025, 0406022034, 0406022038, 0406022017, 0406022014, 0406022011, 0417122078, 0417122081, 0417122084, 0417122087, 0417122090.

View the original FDA recall notice

Why was this product recalled?

The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.

What actions did STERIS CORPORATION take?

Steris issued an URGENT MEDICAL DEVICE RECALL NOTICE dated 06/17/2025 to its consignees on 06/23/2025 disseminated via FedEx. The notice explained the issue, potential risk, and requested the following: User Action: Please ensure the following steps are completed: 1. Review the revised Preventive Maintenance Manual and Interval Based Check List language depending on which HarmonyAIR Surgical Lighting System model(s) are in use at your facility (A-Series and/or E-Series) (please reference Attachments B-E of the letter for maintenance instructions). 2. Immediately complete the maintenance activities as instructed, and then again annually going forward. 3. Complete the Medical Device Recall Response Form included with this letter. 4. Return the completed Response Form via email to: Regulatory_Compliance@STERIS.com, or via fax to 440-392-8963. 5. Distributors: If you have further distributed this product, please identify your Customer(s), and notify them of this recall.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology