Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2192-2025

Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35 Recall

Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35

FDA recall Class II

Ongoing

FDA lists this Class II recall for Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2192-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 21, 2025
FDA report date
Aug 6, 2025
Quantity in commerce
1,064
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35

Affected model & UDI checker

Search the exact identifiers published for Z-2192-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
2.70SPhasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35Check source noticeNot specified
3.20LPhasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35Check source noticeNot specified
4.50Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35Check source noticeNot specified
5.30Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35Check source noticeNot specified
6.35Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35Check source noticeNot specified
FDA code information

UDI-DI: B775270S0, B775320L0, B7754500, B7755300, B7756350. Lot: 241002450, 241002530, 241009450, 241009530, 241016450, 241016530, 241023450, 241023530, 241030450, 241030530, 24112027S, 24121827S, 24122727S, 25012227S, 24082832L, 24102332L, 241227450, 241220635, 250122635, 241227530, 250122530.

View the original FDA recall notice

Why was this product recalled?

Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time.

What actions did PHASOR HEALTH, LLC take?

Starting on 3/21/2025, device correction and removal began and a correction and removal notice was emailed. Customers were asked to do the following: 1) Review inventory (without opening the sterile transparent barrier) to determine if any of the affected lots remain in inventory. If yes, remove those lots from inventory. Do not use drills with reverse battery orientation. 2) If unable to review inventory soon, the Company will come to your site within 30 business days of this letter and inspect your inventory (within 20 minutes total) and replace any defective drills with new verified stock (same catalog number). 3) Contact firm at info@phasorhealth.com to report results of inventory inspection and to facilitate device return. 4) Complete and return the acknowledgement and receipt form via email to info@phasorhealth.com Please contact Firm at info@phasorhealth.com or (832) 982-1234 for more information or questions

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology