Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2198-2025

Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor Recall

Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

FDA recall Class II

Ongoing

FDA lists this Class II recall for Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2198-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 21, 2025
FDA report date
Aug 6, 2025
Quantity in commerce
2160
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Affected model & UDI checker

Search the exact identifiers published for Z-2198-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C20206Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide MonitorCheck source noticeNot specified
Not specifiedPrecision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide MonitorCheck source notice00858481006017
FDA code information

Monitor UDI-DI: 00858481006017. Lot (C20206): 231220

View the original FDA recall notice

Why was this product recalled?

Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.

What actions did Capnia, Inc. take?

Starting on 4/21/2025 Product Advisory and Voluntary Field Action emails were sent to customers who were told affected devices would be replaced. Customers with questions could contact the firm at 904-588-2986

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology