Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor Recall
Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
FDA recall Class II
Ongoing
FDA lists this Class II recall for Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2198-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 21, 2025
- FDA report date
- Aug 6, 2025
- Quantity in commerce
- 2160
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
Affected model & UDI checker
Search the exact identifiers published for Z-2198-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| C20206 | Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor | Check source notice | Not specified |
| Not specified | Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor | Check source notice | 00858481006017 |
FDA code information
Monitor UDI-DI: 00858481006017. Lot (C20206): 231220
Why was this product recalled?
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
What actions did Capnia, Inc. take?
Starting on 4/21/2025 Product Advisory and Voluntary Field Action emails were sent to customers who were told affected devices would be replaced. Customers with questions could contact the firm at 904-588-2986
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2198-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology