Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2205-2026

ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system Recall

ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system

FDA recall Class II

Ongoing

FDA lists this Class II recall for ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2205-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 9, 2026
FDA report date
Jun 10, 2026
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system

Affected model & UDI checker

Search the exact identifiers published for Z-2205-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
10849000ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray systemCheck source noticeNot specified
Not specifiedARTIS pheno, Model 10849000. interventional fluoroscopic x-ray systemCheck source notice4056869046877
FDA code information

UDI: 4056869046877

View the original FDA recall notice

Why was this product recalled?

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

What actions did Siemens Medical Solutions USA, Inc. take?

Distribution of Customer Advisory Notice (CAN) to all affected customers began by Siemens on March 9, 2026.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.