Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2208-2026

Atlan A350XL. Recall

Atlan A350XL. Model Number: 8621600. anesthesia workstation

FDA recall Class I

Ongoing

FDA lists this Class I recall for Atlan A350XL.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2208-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 7, 2026
FDA report date
Jun 10, 2026
Quantity in commerce
394 units
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Atlan A350XL. Model Number: 8621600. anesthesia workstation

Affected model & UDI checker

Search the exact identifiers published for Z-2208-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
8621600Atlan A350XL.Check source noticeNot specified
Not specifiedAtlan A350XL.Check source notice04048675556183
FDA code information

Model Number: 8621600. UDI-DI Number: 04048675556183. Serial Numbers: ASSL-0354, ASSL-0357, ASSL-0360, ASSH-0240, ASSH-0242, ASSH-0243, ASSH-0245, ASSH-0246, ASSH-0247, ASSH-0248, ASSH-0249, ASSH-0250, ASSL-0355, ASSL-0356, ASSL-0358, ASSL-0359, ASSL-0361, ASSH-0239, ASSH-0241, ASSH-0244, ASSH-0251, ASSH-0252, ASSH-0253, ASSH-0254, ASTA-0308, ASTA-0310, ASTA-0312, ASTA-0271, ASTA-0272, ASTA-0273, ASTA-0274, ASTA-0275, ASTA-0276, ASTA-0352, ASTA-0354, ASTA-0364, ASTA-0395, ASTA-0396, ASTA-0397, ASTA-0407, ASSM-0464, ASSM-0465, ASSM-0466, ASSM-0467, ASSM-0468, ASSM-0469, ASSM-0491, ASSM-0492, ASSM-0493, ASSM-0494, ASSM-0495.

View the original FDA recall notice

Why was this product recalled?

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

What actions did Draeger, Inc. take?

Draeger notified consignees of the recall update and expansion on about 05/07/2026 via emailed letter titled Urgent Medical Device Correction Update. Consignees were made aware that due to continuous monitoring, Draeger has determined that an additional number of Atlan devices are affected by the issue and they are replacing ventilator motors. Consignees were instructed that a Draeger Service Representative will contact consignees to arrange a date for the ventilator motor assembly replacement, and until the corrective action is implemented, the device may continue to be used with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Consignees were also instructed to complete and return the provided Medical Device Correction Return Response Acknowledgment and Receipt Form and if products have been made available to third parties, please forward this Urgent Medical Device Correction notice to them.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology