Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2211-2026

Automated Impella Controller (AIC); Product Code: 0042-0000-US; Recall

Automated Impella Controller (AIC); Product Code: 0042-0000-US;

FDA recall Class I

Ongoing

FDA lists this Class I recall for Automated Impella Controller (AIC); Product Code: 0042-0000-US;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2211-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Apr 20, 2026
FDA report date
Jun 3, 2026
Quantity in commerce
91 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Automated Impella Controller (AIC); Product Code: 0042-0000-US;

Affected model & UDI checker

Search the exact identifiers published for Z-2211-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedAutomated Impella Controller (AIC); Product Code: 0042-0000-US;Check source notice00813502010022
FDA code information

Product Code: 0042-0000-US; UDI-DI: 00813502010022; Serial Numbers: IC2449, IC2451, IC2452, IC2453, IC2459, IC2460, IC1096, IC1142, IC1202, IC1203, IC1224, IC1231, IC1238, IC1284, IC1285, IC1390, IC1545, IC1577, IC1665, IC1708, IC1741, IC1852, IC1853, IC2029, IC2071, IC2154, IC2277, IC2278, IC2314, IC2315, IC2368, IC2369, IC2370, IC2442, IC2443, IC2444, IC2461, IC2471, IC2476, IC2477, IC2478, IC2479, IC2575, IC2989, IC3127, IC3128, IC3129, IC3130, IC3166, IC3167, IC3213, IC3223, IC3306, IC3307, IC3308, IC3309, IC3342, IC3398, IC3475, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3674, IC3675, IC3719, IC3720, IC3776, IC3812, IC3887, IC3888, IC4045, IC4060, IC1301, IC1544, IC1572, IC2686, IC2687, IC2848, IC3174, IC3269, IC3272, IC3277, IC3278, IC3281, IC3282, IC3284, IC3294, IC3783;

View the original FDA recall notice

Why was this product recalled?

A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Abiomed, Inc. take?

On April 20, 2026 URGENT MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers via 1-day direct mail (US). OUS notification process will be commensurate with local requirements. Actions to be taken: 1. For the devices identified, the Abiomed servicing team will contact you to coordinate the return of the device(s) to implement necessary changes. To ensure continuity of care, hospital inventory can continue to be used. 2. Upon contact from Abiomed s field servicing team, please work with them to return the identified device(s) for the change to be implemented. Refer to Attachment 2 for list of impacted AIC serial numbers. 3. Forward to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 4. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.