Ivenix Infusion System (IIS), Large Volume Pump Recall
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
FDA recall Class I
Ongoing
FDA lists this Class I recall for Ivenix Infusion System (IIS), Large Volume Pump. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2224-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- May 6, 2026
- FDA report date
- Jun 10, 2026
- Quantity in commerce
- 18,444
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Affected model & UDI checker
Search the exact identifiers published for Z-2224-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| LVP-0004 | Ivenix Infusion System (IIS), Large Volume Pump | Check source notice | Not specified |
| Not specified | Ivenix Infusion System (IIS), Large Volume Pump | Check source notice | 00811505030320 |
FDA code information
Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:
Why was this product recalled?
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
What actions did Fresenius Kabi USA, LLC take?
On May 6, 2026 URGENT - Medical Device Field Correction letters were sent to customers. Required Actions for Healthcare Providers 1. If a unit has been dropped or severely jarred, remove it from service, even if no damage is visible. 2. Once you have removed the pump from service, or if you have questions about whether to do so after a drop, please contact Fresenius Kabi customer support at Ivenix_support@fresenius-kabi.com or (855) 354-6387. 3. Post enclosed Safety Tip Sheet, Proper Handling of the Ivenix Large Volume Pump (LVP) notice within your facilities to ensure that all ancillary staff are informed of this notification and trained on the action to remove any dropped or severely jarred devices from use. 4. Notify other users if applicable. If your facility distributes or transfers devices to other locations, notify all anticipated users of this correction.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2224-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology