Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2231-2025

3M Ranger Irrigation Fluid Warming Set, REF 24750 Recall

3M Ranger Irrigation Fluid Warming Set, REF 24750

FDA recall Class II

Ongoing

FDA lists this Class II recall for 3M Ranger Irrigation Fluid Warming Set, REF 24750. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2231-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 22, 2025
FDA report date
Aug 6, 2025
Quantity in commerce
186,880 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

3M Ranger Irrigation Fluid Warming Set, REF 24750

Affected model & UDI checker

Search the exact identifiers published for Z-2231-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
247503M Ranger Irrigation Fluid Warming Set, REF 24750Check source noticeNot specified
Not specified3M Ranger Irrigation Fluid Warming Set, REF 24750Check source notice50707387792720
Not specified3M Ranger Irrigation Fluid Warming Set, REF 24750Check source notice10707387792722
FDA code information

UDI/DI 50707387792720 (box), 10707387792722 (pouch, )SKU 7100235070, All lots with a manufacturing date after March 2022

View the original FDA recall notice

Why was this product recalled?

Notice was issued to clarify flow rates related to the inlet fluid temperature.

What actions did 3M COMPANY take?

Solventum issued an Urgent Medical Device Correction notice dated 04/21/2025 to its consignees via email and US certified mail. The notice explained the issue, the modification related to the inlet fluid temperature to address the issue, and forthcoming instructions for use updates. All users of the impacted products are being asked to take the following actions. " Ensure this notification is passed on to all those who need to be aware within your organization and to any department where the affected products have been transferred. " Continue to use the products as indicated, considering the updated information provided. " Keep this notice on hand until you are notified that the updated IFUs and operator s manuals are available. " Complete the attached Customer Acknowledgement Form acknowledging that you have read and understood this letter and the updated information provided. " Email the completed form to 3MFieldSafetyAction@montage72.com. For additional questions or need to report an adverse event, contact the Solventum [formerly 3M Health Care] Customer Helpline at 1-800-228-3957, option 6.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA notice posted

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology