COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices are intended for use in providing tracheal access for airway management. Shiley Tracheostomy Tubes are double cannula tracheostomy tubes with reusable inner cannulae (LPC, FEN, CFS, CFN, LGT) and twist-lock connectors. These Shiley Tracheostomy Tubes have a radiopaque, biocompatible outer cannula constructed of polyvinyl chloride. Recall
COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices are intended for use in providing tracheal access for airway management. Shiley Tracheostomy Tubes are double cannula tracheostomy tubes with reusable inner cannulae (LPC, FEN, CFS, CFN, LGT) and twist-lock connectors. These Shiley Tracheostomy Tubes have a radiopaque, biocompatible outer cannula constructed of polyvinyl chloride.
FDA recall Class II
Ongoing
FDA lists this Class II recall for COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices are intended for use in providing tracheal access for airway management. Shiley Tracheostomy Tubes are double cannula tracheostomy tubes with reusable inner cannulae (LPC, FEN, CFS, CFN, LGT) and twist-lock connectors. These Shiley Tracheostomy Tubes have a radiopaque, biocompatible outer cannula constructed of polyvinyl chloride.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2261-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 15, 2026
- FDA report date
- Jun 10, 2026
- Quantity in commerce
- 1,002
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices are intended for use in providing tracheal access for airway management. Shiley Tracheostomy Tubes are double cannula tracheostomy tubes with reusable inner cannulae (LPC, FEN, CFS, CFN, LGT) and twist-lock connectors. These Shiley Tracheostomy Tubes have a radiopaque, biocompatible outer cannula constructed of polyvinyl chloride.
Affected model & UDI checker
Search the exact identifiers published for Z-2261-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 6FEN | COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices are intended for use in providing tracheal access for airway management. Shiley Tracheostomy Tubes are double cannula tracheostomy tubes with reusable inner cannulae (LPC, FEN, CFS, CFN, LGT) and twist-lock connectors. These Shiley Tracheostomy Tubes have a radiopaque, biocompatible outer cannula constructed of polyvinyl chloride. | Check source notice | Not specified |
| Not specified | COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices are intended for use in providing tracheal access for airway management. Shiley Tracheostomy Tubes are double cannula tracheostomy tubes with reusable inner cannulae (LPC, FEN, CFS, CFN, LGT) and twist-lock connectors. These Shiley Tracheostomy Tubes have a radiopaque, biocompatible outer cannula constructed of polyvinyl chloride. | Check source notice | 60884522007028 |
FDA code information
Product Number: 6FEN UDI-DI code: 60884522007028 Lot Number: 25H0868JZX
Why was this product recalled?
Due to customer complaint regarding incorrect display box labeling.
What actions did Covidien LLC take?
On 04/15/2026, the firm sent via mail/email an "Urgent Medical Device Recall" Letter to customers informing them that due to a manufacturing process packaging error, correctly manufactured and labeled individual units of the Covidien Shiley" cuffed fenestrated adult tracheostomy tube with reusable inner cannula were inadvertently placed into incorrect display boxes. A customer report indicating that the product display box is incorrectly labeled as item code 6CFN, which corresponds to cuffless devices. The correct item code is 6FEN, and it appears correctly on all other labeling, including the outer carton labels and device tray label. Customers are instructed to: 1. Quarantine all unused product from the affected lot of Covidien Shiley" cuffed fenestrated adult tracheostomy tube with reusable inner cannula. 2. Return all unused product from the affected lot in their inventory to Medtronic, as described in the Customer Confirmation Form. 3. Share this notice with all relevant personnel within their organization, and with any organization to which the potentially affected product from the specified lot has been transferred or distributed. 4. Complete and return the enclosed Customer Confirmation Form to rs.gmbmitgfca@medtronic.com even if they do not have unused inventory. For questions - contact Medtronic Representative or Customer Service at 800-962-9888.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2261-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology