FDA RECALL RECORD · Z-2274-2026

The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Recall

The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.

FDA recall Class II

Ongoing

FDA lists this Class II recall for The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2274-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 4, 2026
FDA report date
Jun 10, 2026
Quantity in commerce
18,444 installed bases
Distribution
See key details
FDA cause
Device Design

What is being recalled?

The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.

Affected model & UDI checker

Search the exact identifiers published for Z-2274-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
LVP-0004The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS).Check source noticeNot specified
Not specifiedThe Large Volume Pump (LVP) of the Ivenix Infusion System (IIS).Check source notice00811505030320
FDA code information

Model Number: LVP-0004. UDI-DI: 00811505030320

View the original FDA recall notice

Why was this product recalled?

Potential for the Cassette Loading Lever to break.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Fresenius Kabi USA, LLC take?

Fresenius Kabi notified consignees on about 02/04/2026 via letter. Consignees were informed of the lever breakage and that two potential sources of the breakage were: 1. If the Cassette Loading Lever is left open during transport, it can be damaged if it is jarred by another object or surface. 2. Applying force to a locked Loading Lever when it is in the locked position can result in potential damage and/or cause it to break. Consignees were reminded of proper Loading Lever handling: 1. Do not force the Cassette Loading Lever open or closed. The lever does not require significant force to operate, and excessive force can damage the LVP. Excessive force can damage the LVP and prevent use. 2. The Cassette Loading Lever does not require excessive force to load or unload the administration set. 3. If the Cassette Loading Lever does not move with minimal force, do not continue to apply force. Check that the pump is not currently infusing and verify there are no obstructions. 4. When an infusion completes or is paused, the flow dial may take up to 5 seconds to close and unlock the Cassette Loading Lever. Consignees were also reminded of proper transport and storage to ensure the Cassette Loading Lever is fully closed prior to moving, transporting, or storing the pump, and after transport, visually confirm the Cassette Loading Lever is intact and moves freely before clinical use. Additionally, consignees were requested to complete and return the Customer Reply Form provided.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.