The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Recall
The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.
FDA recall Class II
Ongoing
FDA lists this Class II recall for The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS).. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2274-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 4, 2026
- FDA report date
- Jun 10, 2026
- Quantity in commerce
- 18,444 installed bases
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.
Affected model & UDI checker
Search the exact identifiers published for Z-2274-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| LVP-0004 | The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). | Check source notice | Not specified |
| Not specified | The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). | Check source notice | 00811505030320 |
FDA code information
Model Number: LVP-0004. UDI-DI: 00811505030320
Why was this product recalled?
Potential for the Cassette Loading Lever to break.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Fresenius Kabi USA, LLC take?
Fresenius Kabi notified consignees on about 02/04/2026 via letter. Consignees were informed of the lever breakage and that two potential sources of the breakage were: 1. If the Cassette Loading Lever is left open during transport, it can be damaged if it is jarred by another object or surface. 2. Applying force to a locked Loading Lever when it is in the locked position can result in potential damage and/or cause it to break. Consignees were reminded of proper Loading Lever handling: 1. Do not force the Cassette Loading Lever open or closed. The lever does not require significant force to operate, and excessive force can damage the LVP. Excessive force can damage the LVP and prevent use. 2. The Cassette Loading Lever does not require excessive force to load or unload the administration set. 3. If the Cassette Loading Lever does not move with minimal force, do not continue to apply force. Check that the pump is not currently infusing and verify there are no obstructions. 4. When an infusion completes or is paused, the flow dial may take up to 5 seconds to close and unlock the Cassette Loading Lever. Consignees were also reminded of proper transport and storage to ensure the Cassette Loading Lever is fully closed prior to moving, transporting, or storing the pump, and after transport, visually confirm the Cassette Loading Lever is intact and moves freely before clinical use. Additionally, consignees were requested to complete and return the Customer Reply Form provided.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2274-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology