Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2275-2026

PhaseOne Antimicrobial Solution Recall

PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

FDA recall Class II

Ongoing

FDA lists this Class II recall for PhaseOne Antimicrobial Solution. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2275-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 27, 2026
FDA report date
Jun 10, 2026
Quantity in commerce
5904 units
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

Affected model & UDI checker

Search the exact identifiers published for Z-2275-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
15050PhaseOne Antimicrobial SolutionCheck source noticeNot specified
Not specifiedPhaseOne Antimicrobial SolutionCheck source notice00860009268616
FDA code information

UDI:00860009268616 Lot: 25G263

View the original FDA recall notice

Why was this product recalled?

The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Oculus Technologies of Mexico, S.A. de C.V. take?

A MEDICAL DEVICE RECALL notification letter dated 4/27/26 was sent to customers. How to recognize that the device may fail The device may fail if the sprayer component shows internal corrosion, which can affect the appearance of the solution during use. Because the product is packaged in an amber glass bottle, it may be difficult to clearly see changes in the solution inside the container. Therefore, the issue is most likely to be identified during dispensing. Users may recognize this issue by observing a change in color of the solution when sprayed (for example, a light brown or tea-like appearance) or by noticing discoloration within the sprayer mechanism. # Actions to be Taken by the Customer/User As a precaution, any sprayers included with PhaseOne 50 ml bottles should not be used and should be discarded. The solution remains safe and effective for its intended use when it has not been in contact with the sprayer. PhaseOne Health, LLC should inform its customers that product units that have not been assembled with the sprayer may continue to be used safely by carefully pouring the solution directly from the bottle. Product units that have already been assembled with the sprayer should not be used in this configuration and should be removed from use and retained by PhaseOne Health, LLC. Product returned by customers should be held in a designated area and not redistributed or disposed of until further instructions are provided. PhaseOne Health, LLC is requested to retain all returned product and affected inventory at its facility. At this time, Oculus Technologies of Mexico, S.A. de C.V. is not requesting return of product to the manufacturing site. Disposition of returned product, including destruction, will be performed only after alignment with and, if required, approval from the FDA Recall Coordinator. These actions are temporary and are being implemented while a long-term corrective solution is completed. An updated sprayer design intended to prevent

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.