Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28 Recall
Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28
FDA recall Class II
Ongoing
FDA lists this Class II recall for Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2287-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 2, 2025
- FDA report date
- Aug 20, 2025
- Quantity in commerce
- 78 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Lot numbers: V2408xxxx and runs through V2519xxxx UDI: 11-110/00850052017033; 11-310-K/00860005396948; 11-320-K/00860005396986; 11-321-K/00850052017163; 11-310-K-28/00850052017088; 11-320-K-28/00850052017071; 11-321-K-28/00850052017170
Why was this product recalled?
Due to manufacturing issue, neurostimulator device may experience "cable disconnect" error message and interruption to stimulation output.
What actions did Spark Biomedical Inc take?
On July 2, 2025, Spark Biomedical initiated a Urgent Medical Device Recall via phone. During the phone conversation, Spark Biomedical will ask consignees to take the following actions: 1. Spark Biomedical will replace all affected00 units at no cost. Even if the units appear to be working. 2. Current devices may continue to be used until they are replaced. 3. Please perform a check of the cable to patient controller connection before use. If this connection is loose or moves, please discontinue use of that patient controller device and quarantine that device until the replacement arrives. 4. Please notify all appropriate personnel at your facility that need to know about this recall.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2287-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.