Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key Recall
Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2291-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 7, 2023
- FDA report date
- Jun 10, 2026
- Quantity in commerce
- 31 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Affected model & UDI checker
Search the exact identifiers published for Z-2291-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 722079 | Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key | Check source notice | Not specified |
| R2.2.1 | Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key | Check source notice | Not specified |
| R2.2.3 | Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key | Check source notice | Not specified |
| R2.2.5 | Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key | Check source notice | Not specified |
| Not specified | Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key | Check source notice | 00884838099258 |
FDA code information
UDI (01)00884838099258(21)450, (01)00884838099258(21)750, (01)00884838099258(21)409, (01)00884838099258(21)262, (01)00884838099258(21)275, (01)00884838099258(21)277, (01)00884838099258(21)295, (01)00884838099258(21)272, (01)00884838099258(21)261, (01)00884838099258(21)630, (01)00884838099258(21)1003, (01)00884838099258(21)1315, (01)00884838099258(21)381, (01)00884838099258(21)959, (01)00884838085268(21)1013, (01)00884838085268(21)1412, (01)00884838099258(21)232, (01)00884838085268(21)460, (01)00884838085268(21)1387, (01)00884838099258(21)260, (01)00884838099258(21)1019, (01)00884838099258(21)1371, (01)00884838085268(21)40, (01)00884838085268(21)43, (01)00884838085268(21)35, (01)00884838099258(21)1049, (01)00884838099258(21)1484, (01)00884838099258(21)1083, (01)00884838085268(21)703307, (01)00884838099258(21)246, (01)00884838099258(21)475.
Why was this product recalled?
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?
Philips issued an IMPORTANT PRODUCT NOTICE to its consignees on 12/7/2023 via USPS mail. The notice explained the problem with the device and the circumstances under which if could occur and actions to be taken to minimize the effect of the problem. The notice further instructed the user to: - Keep this Important Product Notice with the documentation of the system until Philips corrects your system. - Circulate this notice to all users of the system so that they are aware of the issue. - Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Important Product Notice. For further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2291-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology