FDA RECALL RECORD · Z-2292-2026

Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E Recall

Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E

FDA recall Class II

Ongoing

FDA lists this Class II recall for Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2292-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 23, 2026
FDA report date
Jun 10, 2026
Quantity in commerce
69 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E

Affected model & UDI checker

Search the exact identifiers published for Z-2292-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
IN-28000Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker ECheck source noticeNot specified
FDA code information

All Lots/serial Numbers; UDI: IN-28000: 00810180350020; IN-10500: (01)00810180350686; IN-10600 (01)00810180350693; IN-16400:(01)00810180350525; IN-16100: (01)00810180350563; IN-28050: 00810180350051

View the original FDA recall notice

Why was this product recalled?

The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.

What actions did Kico Knee Innovation Company take?

On April 27, 2026 Kico Knee Innovation issued a Urgent: Medical Device Recall to affected consignees via E-Mail. Kico ask consignees to take the following actions: 1. Instruments can continue to be used as is, given none of the failure modes are exhibited. 2. This notice needs to be passed on to all those who need to be aware within your organization or where the potentially affected devices have been transferred. 3. Please retain a copy of this letter, place a copy where devices are stored, and keep a copy of the acknowledgement form for your files. 4. Complete the attached Acknowledgement Form immediately and return it email to Alexander.Borchardt@enovis.com 5. Instruments should always be visually inspected prior to use as stated in the Instructions for Use.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology