Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E Recall
Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E
FDA recall Class II
Ongoing
FDA lists this Class II recall for Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2292-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 23, 2026
- FDA report date
- Jun 10, 2026
- Quantity in commerce
- 69 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E
Affected model & UDI checker
Search the exact identifiers published for Z-2292-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| IN-28000 | Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E | Check source notice | Not specified |
FDA code information
All Lots/serial Numbers; UDI: IN-28000: 00810180350020; IN-10500: (01)00810180350686; IN-10600 (01)00810180350693; IN-16400:(01)00810180350525; IN-16100: (01)00810180350563; IN-28050: 00810180350051
Why was this product recalled?
The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
What actions did Kico Knee Innovation Company take?
On April 27, 2026 Kico Knee Innovation issued a Urgent: Medical Device Recall to affected consignees via E-Mail. Kico ask consignees to take the following actions: 1. Instruments can continue to be used as is, given none of the failure modes are exhibited. 2. This notice needs to be passed on to all those who need to be aware within your organization or where the potentially affected devices have been transferred. 3. Please retain a copy of this letter, place a copy where devices are stored, and keep a copy of the acknowledgement form for your files. 4. Complete the attached Acknowledgement Form immediately and return it email to Alexander.Borchardt@enovis.com 5. Instruments should always be visually inspected prior to use as stated in the Instructions for Use.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2292-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology