Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Recall
Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2296-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 24, 2026
- FDA report date
- Jun 10, 2026
- Quantity in commerce
- 11,300 units (1650 US, 9650 OUS)
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Catalog number: AVA 500 MPAT (Part 2 of 2) Affected material numbers: 87629007, 60729458, 86566648; Affected batch numbers: 251702, 252302, 252502, 252602, 252702, 252802, 253304, 253702, 253704, 253804, 254202, 254304, 254704, 254802;
Why was this product recalled?
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
What actions did Bayer Medical Care, Inc. take?
On April 24, 2026, URGENT MEDICAL DEVICE RECALL letters were sent to customers. Customers were instructed to immediately halt the use of the affected product and review inventory for impacted product and quarantine. A Return Goods Authorization will be sent to return any remaining affected inventory. Place the affected product in a box and label the outside of the container with the RGA number provided and RECALL ID: SA-2026-RAD-03 in large, bold writing. Use the shipping label provided via email when you receive your RGA number. Please return the affected product as soon as possible. Please distribute this letter to other appropriate departments and personnel within your facility who may need awareness of this recall.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2296-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.