3mensio Workstation (Vascular Fenestrated) software Recall
3mensio Workstation (Vascular Fenestrated) software
FDA recall Class II
Ongoing
FDA lists this Class II recall for 3mensio Workstation (Vascular Fenestrated) software. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2297-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 14, 2025
- FDA report date
- Aug 20, 2025
- Quantity in commerce
- 501
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
3mensio Workstation (Vascular Fenestrated) software
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Software Version/UDI/Software Release Date: 10.6/(01)08056304455505(11)240325(8012)10.6/05-Apr-24, 10.7/(01)08056304455925(11)250423(8012)10.7/29-Apr-25, Service Packs/UDI: 10.6 SP1/(01)08056304455505(11)240411(8012)10.6 SP1/23-Apr-24, 10.6 SP2/(01)08056304455505(11)240610(8012)10.6 SP2/17-Jun-24, 10.6 SP3/(01)08056304455505(11)240926(8012)10.6 SP3/03-Oct-24, 10.6 SP4/(01)08056304455505(11)250109(8012)10.6 SP4/20-Jan-25, 10.7 SP1/(01)08056304455925(11)250523(8012)10.7 SP1/28-May-25
Why was this product recalled?
When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage
What actions did PIE Medical Imaging B.V. take?
On 7/14/2025, correction notices were mailed and emailed to customers who were asked to do the following: When using software version 10.6, in case you open a session state with saved clock measurements and there is a need to adjust the 12h position for the clock positions, make sure to update all individual clock measurements in the perpendicular plane by the following steps: 1. Click on the clock position. 2. Update the 12h pointer of the clock in the perpendicular viewport. 3. Perform step 1 and 2 for all clock measurements in the analysis. After updating all 12h pointers, the information shown in the fenestration diagram is correct and functionality is restored. - Pass and post this notice for all who need to be aware within your organization or to any organization where the potentially affected product is in use. - Complete and return the Customer Acknowledgement Form via email to vigilance@pie.nl Questions can be sent to the above email. Firm will reach out to all customers to provide a link for the software installation.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2297-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology