MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph Recall
MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph
FDA recall Class II
Ongoing
FDA lists this Class II recall for MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2306-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 21, 2026
- FDA report date
- Jun 10, 2026
- Quantity in commerce
- 16
- Distribution
- See key details
- FDA cause
- Software Manufacturing/Software Deployment
What is being recalled?
MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph
Affected model & UDI checker
Search the exact identifiers published for Z-2306-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph | Check source notice | 00195278276070 |
| Not specified | MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph | Check source notice | 00195278276124 |
FDA code information
Software version 3.00 855001-001, UDI-DI 00195278276070 8855002-001, UDI-DI 00195278276124
Why was this product recalled?
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
What actions did GE Medical Systems China Co., Ltd. take?
Firm notified consignees beginning April 21, 2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION". Customers were given instructions on how to check their software version. If a customer's software version begins with 3.00, the customer should not use the system until a GE HealthCare representative has updated the device with the latest licensed software version.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2306-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026