Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2308-2026

TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Recall

TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis

FDA recall Class II

Ongoing

FDA lists this Class II recall for TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2308-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 3, 2026
FDA report date
Jun 10, 2026
Quantity in commerce
7
Distribution
See key details
FDA cause
Process control

What is being recalled?

TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis

Affected model & UDI checker

Search the exact identifiers published for Z-2308-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
409-375TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM LCheck source noticeNot specified
Not specifiedTITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM LCheck source notice00844505000772
FDA code information

Model/Catalog Number: 409-375 UDI code: (01)00844505000772(10)122927(17)310301 Lot Number: 122927

View the original FDA recall notice

Why was this product recalled?

Due to incorrect functional length on device labeling.

What actions did Grace Medical, Inc. take?

On 04/03/2026, the firm emailed a "RECALL NOTICE" to it single customer informing them that the firm determined the pistons were manufactured with an incorrect functional length. Devices are labeled as 3.75 mm but have an actual functional length of 4.00 mm. The customer was instructed to: " Immediately discontinue use and quarantine affected product " Return affected product using the provided RMA " If any devices have been used, provide the date(s) of surgery on the response form " Complete and return the enclosed Recall Response Form For questions - contact Grace Medical at 901-386-0990 or Thomas.fearnley@gracemedical.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology
Z-2308-2026: TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L recall | MedtechLedger