Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Recall
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2339-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 25, 2025
- FDA report date
- Aug 27, 2025
- Quantity in commerce
- 434 units
- Distribution
- See key details
- FDA cause
- Component change control
What is being recalled?
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
Affected model & UDI checker
Search the exact identifiers published for Z-2339-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| WA91307C | Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. | Check source notice | Not specified |
| Not specified | Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. | Check source notice | 04042761086349 |
FDA code information
Model Number: WA91307C. Lot Code: ESG-410 Legacy WA91307C. UDI-DI: 04042761086349. Serial Numbers: 101004, 101005, 101006, 101007, 101008, 101009, 101010, 101011, 101012, 101013, 101014, 101015, 101016, 101017, 101018, 101019, 101020, 101021, 101022, 101023, 101024, 101025, 101029, 101030, 101031, 101032, 101033, 101034, 101035, 101036, 101037, 101038, 101039, 101040, 101041, 101042, 101043, 101044, 101045, 101046, 101047, 101048, 101049, 101050, 101051, 101052, 101053, 101054, 101055, 101056, 101057, 101058, 101059, 101060, 101061, 101062, 101063, 101064, 101066, 101067, 101068, 101069, 101070, 101071, 101072, 101073, 101074, 101238, 101239, 101240, 101241, 101242, 101243, 101244, 101245, 101246, 101247, 101250, 101251, 101252, 101253, 101254, 101255, 101256, 101257, 101258, 101259, 101260, 101261, 101262, 101263, 101264, 101265, 101266, 101267, 101268, 101269, 101270, 101271, 101272, 101273, 101274, 101275, 101276, 101277, 101278, 101279, 101280, 101281, 101282, 101283, 101284, 101285, 101286, 101287, 101288, 101289, 101290, 101291, 101292, 101293, 101294, 101295, 101296, 101297, 101298, 101299, 101300, 101301, 101302, 101333, 101334, 101335, 101336, 101337, 101338, 101339, 101340, 101341, 101342, 101370, 101371, 101372, 101373, 101374, 101375, 101376, 101377, 101378, 101379, 101395, 101396, 101397, 101398, 101399, 101400, 101401, 101402, 101403, 101432, 101433, 101434, 101435, 101436, 101437, 101438, 101439, 101440, 101441, 101613, 101633, 101634, 101635, 101636, 101637, 101638, 101684, 101685, 101686, 101687, 101688, 101689, 101691, 101690, 101692, 101693, 101694, 101695, 101743, 101744, 101745, 101746, 101747, 101748, 101749, 101750, 101751, 101752, 101753, 101754, 101755, 101756, 101757, 101758, 101759, 101760, 101761, 101762, 101763, 101764, 101765, 101766, 101767, 101768, 101769, 101770, 101771, 101772, 101773, 101774, 101775, 101776, 101777, 101778, 101779, 101780, 101781, 101782, 101783, 101784, 101785, 101786, 101787, 101788, 101789, 101958, 101959, 101960, 101961, 101962, 101963, 101964, 101965, 101966, 101967, 101968, 101969, 101970, 101971, 102017, 102018, 102019, 102020, 102021, 102024, 102025, 102026, 102027, 102028, 102055, 102056, 102057, 102058, 102059, 102060, 102061, 102062, 102063, 102064, 102065, 102066, 102067, 102068, 102069, 102284, 102285, 102286, 102287, 102288, 102289, 102290, 102291, 102292, 102293, 102294, 102295, 102296, 102297, 102298, 102299, 102300, 102301, 102302, 102303, 102304, 102305, 102306, 102307, 102308, 102309, 102310, 102311, 102312, 102313, 102314, 102315, 102316, 102317, 102318, 102319, 102320, 102321, 102323, 102325, 102326, 102327, 102328, 102329, 102330, 102331, 102332, 102333, 102334, 102335, 102362, 102363, 102364, 102366, 102367, 102368, 102369, 102370, 102371, 102372, 102373, 102391, 102392, 102393, 102394, 102395, 102396, 102397, 102399, 102400, 102401, 102402, 102403, 102404, 102405, 102406, 102407, 102408, 102409, 102410, 102411, 102412, 102413, 102414, 102415, 102416, 102417, 102418, 102419, 102420, 102421, 102422, 102423, 102424, 102425, 102426, 102428, 102429, 102430, 102431, 102432, 102433, 102434, 102435, 102436, 102437, 102438, 102439, 102440, 102441, 102442, 102443, 102444, 102445, 102497, 102498, 102499, 102500, 102501, 102502, 102503, 102504, 102505, 102506, 102507, 102508, 102509, 102510, 102511, 102512, 102513, 102515, 102516, 102517, 102518, 102519, 102520, 102521, 102522, 102523, 102802, 102807, 102808, 102809, 102810, 102814, 102818, 102819, 102820, 102821, 102822, 102823, 102828, 102829, 102830, 102831, 102832, 102834, 102836, 102837.
Why was this product recalled?
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Olympus Corporation of the Americas take?
Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further distributed, and to acknowledge receipt of the letter through the Olympus web portal.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2339-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology