Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual Recall
Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
FDA recall Class II
Ongoing
FDA lists this Class II recall for Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2340-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 8, 2025
- FDA report date
- Aug 27, 2025
- Quantity in commerce
- 24,583
- Distribution
- See key details
- FDA cause
- Component change control
What is being recalled?
Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
Affected model & UDI checker
Search the exact identifiers published for Z-2340-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| M365DB4600C0 | Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual | Check source notice | Not specified |
| M365DB4605C0 | Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual | Check source notice | Not specified |
FDA code information
All lots. Surgical Implant Manual version REF/UDI-DI: M365DB4600C0/08714729820802, M365DB4605C0/08714729820819
Why was this product recalled?
Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.
What actions did Boston Scientific Neuromodulation Corporation take?
On 7/8/2025, correction notices were mailed, emailed, or delivered to healthcare professionals who were asked to do the following: 1. Review the updates to the Surgical Implant Manual stated in this notice. These updates will be found in the IFU once implemented. 2. If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. 3. If you are a distributor, this notification must be forwarded to your customers to ensure this notification is carried out to the end-user level. 4. To provide awareness of this information, share this letter with any other clinicians in your hospital who use these devices. 5. If users encounter resistance or difficulties while closing the Retaining Clip locking mechanism, refer to the Surgical Implant Manual updates provided. 6. Maintain a copy of this letter in your facility s records. 7. Complete and return the acknowledgement form via email to BSCFieldActionCenter@bsci.com The Surgical Implant Manual has updates that include: - If you have difficulty locking the Slider, you may need to remove the Lead Stylet before locking the Slider in place, or replace the Retaining Clip using another Burr Hole Cover Kit or Burr Hole Cover Spares Kit. - You may confirm the Slider is fully locked. - Verify that the DBS Lead has not moved from the desired location (e.g. by intraoperative imaging). Report all device-related incidents or quality concerns experienced with the use of these devices to firm at BSN.ComplaintCallCenter@bsci.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2340-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026