Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2341-2025

GE HealthCare OEC 9800. Recall

GE HealthCare OEC 9800.

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE HealthCare OEC 9800.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2341-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 2, 2025
FDA report date
Sep 3, 2025
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

GE HealthCare OEC 9800.

Affected model & UDI checker

Search the exact identifiers published for Z-2341-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedGE HealthCare OEC 9800.Check source notice00840682114349
FDA code information

GTIN: 00840682114349

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

What actions did GE Medical Systems, LLC take?

GE HealthCare will send a customer letter providing safety instructions as an immediate mitigation, to communicate the risk of additional radiation and that the equipment should only be used by personnel properly trained and authorized in the use of ionizing radiation. A user should observe proper safety practices during equipment operation as required by the operator manual. GE HealthCare will correct all affected OEC 9800 and OEC 9900 Elite systems (28 units) by visually inspecting for the presence of lead shield tape on the collimator cover and replacing the collimator cover if it is found non-compliant, at no cost to the customer.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.