FDA RECALL RECORD · Z-2345-2025

Dynex Agility Sample Tips; Recall

Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Dynex Agility Sample Tips;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2345-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 16, 2025
FDA report date
Aug 27, 2025
Quantity in commerce
237 units
Distribution
See key details
FDA cause
Labeling Change Control

What is being recalled?

Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system.

Affected model & UDI checker

Search the exact identifiers published for Z-2345-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
67910Dynex Agility Sample Tips;Check source noticeNot specified
Not specifiedDynex Agility Sample Tips;Check source notice5060456180072
FDA code information

Model Number: 67910; UDI-DI: 5060456180072; Lot Number: 863366;

View the original FDA recall notice

Why was this product recalled?

The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.

What actions did Dynex Technologies, Inc. take?

On July 16, 2025, URGENT FIELD SAFETY NOTICE letters were emailed to customers. Action to Be Taken by the User: - Upon receipt, Quarantine affected devices and return or destroy them. - Customers have been provided a Field Safety Notice Customer Reply Form. Please complete this form, sign it and return to Dynex as soon as possible to acknowledge that you have received this information and have destroyed or returned to Dynex the impacted 67910 Agility Sample Tips Lot 863366. Transmission of this Field Safety Notice This notice needs to be passed on all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred. (As appropriate) Please transfer this notice to other organisations on which this action has an impact. (As appropriate) Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents to the manufacturer, distributor or local representative, and the national Competent Authority if appropriate, as this provides important feedback..*

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA notice posted

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.