SD LTM 64 PLUS, EEG Amplifier/recorder Recall
SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02
FDA recall Class II
Ongoing
FDA lists this Class II recall for SD LTM 64 PLUS, EEG Amplifier/recorder. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2353-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 4, 2026
- FDA report date
- Jun 17, 2026
- Quantity in commerce
- 207 units
- Distribution
- See key details
- FDA cause
- Software Design Change
What is being recalled?
SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02
Affected model & UDI checker
Search the exact identifiers published for Z-2353-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | SD LTM 64 PLUS, EEG Amplifier/recorder | Check source notice | 8033928120897 |
FDA code information
UDI-DI: 8033928120897; Lot Code: Model No SD LTM 64 PLUS; UDI-DI 8033928120897; Serial Numbers BAA-0051/01-19, BAA-0069/02-19, BAA-0070/02-19, BAA-0071/02-19, BAA-0082/02-19, BAA-0083/02-19, BAA-0084/02-19, BAA-0085/02-19, BAA-0086/02-19, BAA-0087/02-19, BAA-0088/02-19, BAA-0089/02-19, BAA-0090/02-19, BAA-0091/02-19, BAA-0092/02-19, BAA-0093/02-19, BAA-0095/02-19, BAA-0096/02-19, BAA-0098/02-19, BAA-0104/02-19, BAA-0105/02-19, BAA-0106/02-19, BAA-0107/02-19, BAA-0108/02-19, BAA-0109/02-19, BAA-0110/02-19, BAA-0111/02-19, BAA-0112/02-19, BAA-0113/02-19, BAA-0115/02-19, BAA-0116/02-19, BAA-0117/02-19, BAA-0118/02-19, BAA-0119/02-19, BAA-0121/02-19, BAA-0122/02-19, BAA-0142/03-19, BAA-0155/03-20, BAA-0156/03-20, BAA-0191/04-20, BAA-0192/04-20, BAA-0193/04-20, BAA-0194/04-20, BAA-0286/04-21, BAA-0287/04-21, BAA-0288/04-21, BAA-0289/04-21, BAA-0329/04-22
Why was this product recalled?
Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Micromed S.p.A. take?
An Urgent Medical Device Recall notification letter dated 4/27/26 was sent to customers. ACTION: Natus is requesting you to check your SD LTM 64 PLUS stock and identify whether the devices are currently running firmware version 2021.02, 2022.01 or 2022.02. The information is displayed on the device screen when the device is switched on if you keep the switch on button pressed. " Immediately cease use of the affected product in configurations with multiple amplifiers (i.e, systems with 128- or 256-channels) until you receive the instructions from Natus to return it or have it serviced locally. " Please complete the enclosed form and return to FCA@Natus.com. If you have affected devices with firmware 2021.02, 2022.01 or 2022.02, you will receive detailed instructions from FCA@Natus.com. " Adverse reactions or quality problems experienced with the use of the part should be reported to Natus and the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report Online at www.fda.gov/medwatch/report.htm. " Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178. " In the event you have affected product, once you provide the completed Customer Reply Form that is listed below to FCA@natus.com, Technical Service will be in contact with you to arrange for local servicing or return the device(s) for service. RESOLUTION: " Issue is resolved by updating the product firmware to version 2023.01 or later. " Natus has identified the root cause of this issue and has implemented corrective actions to prevent future recurrence. If you have any questions regarding this notice, please contact FCA@Natus.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2353-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology