Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2400-2026

Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28 Recall

Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28

FDA recall Class II

Ongoing

FDA lists this Class II recall for Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2400-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 27, 2026
FDA report date
Jun 17, 2026
Quantity in commerce
500
Distribution
See key details
FDA cause
Labeling mix-ups

What is being recalled?

Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28

Affected model & UDI checker

Search the exact identifiers published for Z-2400-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
504600-28Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28Check source noticeNot specified
Not specifiedAdvance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28Check source notice00371347008104
FDA code information

Model No. 504600-28; UDI-DI 00371347008104; Lot Number 2026-0204

View the original FDA recall notice

Why was this product recalled?

Some units of Aluminum Chloride Gel contain the package insert for Ferric Sulfate Gel. The incorrect insert misidentifies the product and contains different first aid instructions, most critically in that the Ferric Sulfate insert instructs users to induce vomiting in the case of ingestion, while the correct Aluminum Chloride insert instructs users not to induce vomiting.

What actions did Inter-Med Llc take?

On 5/27/2026, firm notified consignees via email. Consignees were instructed to return all units from Lot 2026-0204 to Inter-Med, Inc. Inter-Med, Inc. will provide prepaid return shipping labels. All returned product will be replaced with correctly packaged product at no charge, or a credit may be issued at the customer's preference. Customers must also notify any downstream dental practices or end users who received product from this lot and instruct them to return product.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.