Bicarby Dialysate; Recall
Bicarby Dialysate; Model number: RFP-400-G;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Bicarby Dialysate;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2401-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 7, 2026
- FDA report date
- Jun 17, 2026
- Quantity in commerce
- 4752
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Bicarby Dialysate; Model number: RFP-400-G;
Affected model & UDI checker
Search the exact identifiers published for Z-2401-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| RFP-400-G | Bicarby Dialysate; | Check source notice | Not specified |
| Not specified | Bicarby Dialysate; | Check source notice | 00840861102839 |
FDA code information
Model number: RFP-400-G; UDI-DI: 00840861102839; Lot codes: 25SG06026, 25SG06027;
Why was this product recalled?
There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Fresenius Medical Care Holdings, Inc. take?
On 5/7/2026 an URGENT MEDICAL DEVICE RECALL letters were sent to customers. At this time, FMCRTG is providing this recall notice to alert you of the potential risks associated with the labeling error, and to ask that you identify, segregate, and discontinue use of affected lots, 25SG06026 and 25SG06027, of RFP-400-G Bicarby Dialysate. FMCRTG is requiring the return of the affected Bicarby Dialysate lots. FMCRTG will replace all remaining stock of affected product lots. For product technical questions or concerns, please contact Technical Support at 1-800-323-2732 or Customer Support at 1-800-323-5188 between 8:30AM 4:30PM Eastern Daylight Time. For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855-616-2309.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2401-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology