Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software Recall
Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software
FDA recall Class II
Ongoing
FDA lists this Class II recall for Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2405-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 12, 2026
- FDA report date
- Jun 17, 2026
- Quantity in commerce
- 11182 programs
- Distribution
- See key details
- FDA cause
- Software Design Change
What is being recalled?
Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software
Affected model & UDI checker
Search the exact identifiers published for Z-2405-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 3650 | Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software | Check source notice | Not specified |
| 3330 | Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software | Check source notice | Not specified |
| Not specified | Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software | Check source notice | 05414734509725 |
FDA code information
Programmer Model 3650; Software Model 3330; GTIN 05414734509725; Software Versions 25.0.1 - 28.7.1 in the United States (varies based on region)
Why was this product recalled?
Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.
What actions did ABBOTT MEDICAL take?
On May 12 Abbott issued a Urgent Medical Device Recall notification via sales representative. Abbott ask consignees to take the following actions: Abbott has developed updated Merlin PCS 3650 programmer software (v28.9.1 rev 1 or higher) that resolves this behavior. Beginning with this notification, your Abbott Representative will upgrade programmer software to this software version . Abbott's service representative will proved a attestation to acknowledge receipt of the notification and software update. Please share this notification with others in your organization, as appropriate.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2405-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology