Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software Recall

Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software

FDA recall Class II

Ongoing

FDA lists this Class II recall for Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2405-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 12, 2026
FDA report date
Jun 17, 2026
Quantity in commerce
11182 programs
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software

Affected model & UDI checker

Search the exact identifiers published for Z-2405-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
3650Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 softwareCheck source noticeNot specified
3330Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 softwareCheck source noticeNot specified
Not specifiedMerlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 softwareCheck source notice05414734509725
FDA code information

Programmer Model 3650; Software Model 3330; GTIN 05414734509725; Software Versions 25.0.1 - 28.7.1 in the United States (varies based on region)

View the original FDA recall notice

Why was this product recalled?

Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.

What actions did ABBOTT MEDICAL take?

On May 12 Abbott issued a Urgent Medical Device Recall notification via sales representative. Abbott ask consignees to take the following actions: Abbott has developed updated Merlin PCS 3650 programmer software (v28.9.1 rev 1 or higher) that resolves this behavior. Beginning with this notification, your Abbott Representative will upgrade programmer software to this software version . Abbott's service representative will proved a attestation to acknowledge receipt of the notification and software update. Please share this notification with others in your organization, as appropriate.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology