Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2415-2025

Spectrum IQ Infusion Pump, Product Code 3570009 Recall

Spectrum IQ Infusion Pump, Product Code 3570009

FDA recall Class II

Ongoing

FDA lists this Class II recall for Spectrum IQ Infusion Pump, Product Code 3570009. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2415-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 14, 2025
FDA report date
Sep 3, 2025
Quantity in commerce
6 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Spectrum IQ Infusion Pump, Product Code 3570009

Affected model & UDI checker

Search the exact identifiers published for Z-2415-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSpectrum IQ Infusion Pump, Product Code 3570009Check source notice00085412610900
FDA code information

UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271

View the original FDA recall notice

Why was this product recalled?

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Baxter Healthcare Corporation take?

Customers were contacted via telephone beginning 7/14/25. The included information about the product, the issue and the action to be taken by the customers. Baxter Healthcare is requesting the return of the impacted product to perform all tests required and confirm that the product perform as expected. If you have any questions, phone (Toll Free): 800-422-9837 or 847-948-4770; hours of operation: Monday through Friday from 8:00 am to 5:00 pm CST

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology