Medline convenience kits: DAVINCI MAIN DYNJ908031D Recall
Medline convenience kits: DAVINCI MAIN DYNJ908031D
FDA recall Class II
Ongoing
FDA lists this Class II recall for Medline convenience kits: DAVINCI MAIN DYNJ908031D. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2434-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 27, 2026
- FDA report date
- Jun 24, 2026
- Quantity in commerce
- 49,654 kits total
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Medline convenience kits: DAVINCI MAIN DYNJ908031D
Affected model & UDI checker
Search the exact identifiers published for Z-2434-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Medline convenience kits: DAVINCI MAIN DYNJ908031D | Check source notice | 10198459632297 |
FDA code information
DYNJ908031D UDI-DI 10198459632297 lot 26AMB067
Why was this product recalled?
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
What actions did Medline Industries, LP take?
Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2434-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology