FDA RECALL RECORD · Z-2438-2025

PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724 Recall

PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724

FDA recall Class II

Ongoing

FDA lists this Class II recall for PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2438-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 28, 2025
FDA report date
Sep 3, 2025
Quantity in commerce
2449 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724

Affected model & UDI checker

Search the exact identifiers published for Z-2438-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedPrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724Check source notice07332414126018
FDA code information

UDI 07332414126018, All serial numbers

View the original FDA recall notice

Why was this product recalled?

Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD Self-Test Failure

What actions did Vantive US Healthcare, LLC take?

Vantive issued an Urgent Medical Device Correction notice to it consignees on 0728/2025 via USPS first class mail. The notice explained the issues with the devices, hazard involved, and requested the following: Actions to be Taken by Customers 1. Operators may continue to safely use the PrisMax Systems and TherMax Blood Warmer Units per the associated Operator s Manuals and on-screen instructions. If further assistance is required, please contact your local Vantive account representative. Please ensure that all operators of these devices are made aware of this Urgent Medical Device Correction notification. Vantive recommends that a copy of this notification is posted within the appropriate location of your facility. 2. As corrections become available, a local Vantive representative will contact your facility to determine the correction plan and schedule the associated correction(s) for impacted devices. 3. If you received this communication directly from Vantive, complete the enclosed Vantive Customer Reply Form and return it to Vantive by scanning and e-mailing it to FieldActions@vantive.com. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. If you do not return the customer reply form, you will receive an email and/or phone call from Vantive to confirm your receipt of this notification. 4. If you purchased this product from a distributor, please note that responding using the Vantive Customer Reply Form is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology