Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2456-2025

IDM-MICRO-R. Electrical wheelchair component. Recall

IDM-MICRO-R. Electrical wheelchair component.

FDA recall Class I

Ongoing

FDA lists this Class I recall for IDM-MICRO-R. Electrical wheelchair component.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2456-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jul 10, 2025
FDA report date
Sep 10, 2025
Quantity in commerce
15 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

IDM-MICRO-R. Electrical wheelchair component.

Affected model & UDI checker

Search the exact identifiers published for Z-2456-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PRSPS0014IDM-MICRO-R. Electrical wheelchair component.Check source noticeNot specified
Not specifiedIDM-MICRO-R. Electrical wheelchair component.Check source notice05407008320836
FDA code information

Model No. PRSPS0014; UDI-DI: 05407008320836; Serial No. 1000 to 1690.

View the original FDA recall notice

Why was this product recalled?

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did mo-Vis BVBA take?

An URGENT: MEDICAL DEVICE RECALL R-NET JOYSTICKS notification dated 07/10/2025 was sent to consignees instructing them to upgrade affected devices according to the provided Annex 2 or by emailing contact@mo-vis.com to coordinate return. The software upgrade of affected devices should be conducted during the next scheduled annual maintenance of the joystick. These actions should be verified. Questions can be directed via email to contact@mo-vis.com or by phone at +32 9 335 28 60.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology