Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2457-2026

Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP. Recall

Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2457-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 18, 2026
FDA report date
Jun 24, 2026
Quantity in commerce
10 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.

Affected model & UDI checker

Search the exact identifiers published for Z-2457-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedImpella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.Check source notice00813502012279
FDA code information

GTIN: 00813502012279. Product Codes (Serial Number): 1000080 (677223, 672986, 673252, 644591, 645428, 644314, 613525), 1000851 (636173), 0048-0014 (499250), 0048-0024-JP (661216).

View the original FDA recall notice

Why was this product recalled?

Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Abiomed, Inc. take?

Johnson & Johnson MedTech/Abiomed Issued an URGETN MEDICAL DEVICE RECALL (REMOVAL) notice to its consignees on 05/18/2026 via UPS 1-day air delivery. The notice explained the reason for notification and potential patient impact Consignees were instructed to review all Impella CP Sets within inventory and quarantine affected units on hand, return all affected units using the provided return shipment label from Sedgwick, and complete and return the provided business response form. Upon receipt of returned product and completed BRF, Abiomed will provide a copy of a credit memo for the amount returned. Consignees were also instructed to forward the notification to anyone in each facility that needs to be informed and to post a copy of the notification in a visible area for awareness. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.