Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP. Recall
Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2457-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- May 18, 2026
- FDA report date
- Jun 24, 2026
- Quantity in commerce
- 10 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
Affected model & UDI checker
Search the exact identifiers published for Z-2457-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP. | Check source notice | 00813502012279 |
FDA code information
GTIN: 00813502012279. Product Codes (Serial Number): 1000080 (677223, 672986, 673252, 644591, 645428, 644314, 613525), 1000851 (636173), 0048-0014 (499250), 0048-0024-JP (661216).
Why was this product recalled?
Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Abiomed, Inc. take?
Johnson & Johnson MedTech/Abiomed Issued an URGETN MEDICAL DEVICE RECALL (REMOVAL) notice to its consignees on 05/18/2026 via UPS 1-day air delivery. The notice explained the reason for notification and potential patient impact Consignees were instructed to review all Impella CP Sets within inventory and quarantine affected units on hand, return all affected units using the provided return shipment label from Sedgwick, and complete and return the provided business response form. Upon receipt of returned product and completed BRF, Abiomed will provide a copy of a credit memo for the amount returned. Consignees were also instructed to forward the notification to anyone in each facility that needs to be informed and to post a copy of the notification in a visible area for awareness. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2457-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology