FDA RECALL RECORD · Z-2463-2026

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. Recall

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2463-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 29, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
27,270 pcs
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Affected model & UDI checker

Search the exact identifiers published for Z-2463-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
260127Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.Check source noticeNot specified
260128Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.Check source noticeNot specified
260167Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.Check source noticeNot specified
HAMILTON-T1Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.Check source noticeNot specified
FDA code information

REF/UDI-DI: 260127/07630002802956, 17630002802953; 260128/07630002802963, 17630002802960; 260167/07630002802970, 17630002802977; 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.)

View the original FDA recall notice

Why was this product recalled?

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Hamilton Medical AG take?

On 5/29/2026, correction notices were emailed to customers who were asked to do the following: 1) Inform all potential users of affected devices within your facility about this issue. 2) Inform all affected end users and/or sub-distributors and provide them with this notice. 5) Post this notice to ensure all personnel are informed. 6) Complete and return Facility Reply Form via email to complaints@hamiltonmedical.com Customers with questions can contact the firm at the email above. Initially customers were informed to continue use and perform actions to mitigate risk, but later the updated notice called for affected device removal. On 7/8/2026, updated recall notices were emailed and mailed to healthcare facilities, physicians and end users who were informed of the following: 1) Discontinue use and quarantine affected devices. 2) Do not follow previous instructions for product modification. 3) Inform all affected end users and/or sub-distributors and provide them with this notice. 4) Contact firm if you have affected product in inventory. - Discard any affected product(s) and confirm destruction. Or - Contact Hamilton Medical for a return material authorization number and return label. -Contact your sales representative to acquire replacement product. 5) Complete and return the Facility Reply Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.