Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. Recall
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2463-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- May 29, 2026
- FDA report date
- Jul 8, 2026
- Quantity in commerce
- 27,270 pcs
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Affected model & UDI checker
Search the exact identifiers published for Z-2463-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 260127 | Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. | Check source notice | Not specified |
| 260128 | Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. | Check source notice | Not specified |
| 260167 | Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. | Check source notice | Not specified |
| HAMILTON-T1 | Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. | Check source notice | Not specified |
FDA code information
REF/UDI-DI: 260127/07630002802956, 17630002802953; 260128/07630002802963, 17630002802960; 260167/07630002802970, 17630002802977; 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.)
Why was this product recalled?
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Hamilton Medical AG take?
On 5/29/2026, correction notices were emailed to customers who were asked to do the following: 1) Inform all potential users of affected devices within your facility about this issue. 2) Inform all affected end users and/or sub-distributors and provide them with this notice. 5) Post this notice to ensure all personnel are informed. 6) Complete and return Facility Reply Form via email to complaints@hamiltonmedical.com Customers with questions can contact the firm at the email above. Initially customers were informed to continue use and perform actions to mitigate risk, but later the updated notice called for affected device removal. On 7/8/2026, updated recall notices were emailed and mailed to healthcare facilities, physicians and end users who were informed of the following: 1) Discontinue use and quarantine affected devices. 2) Do not follow previous instructions for product modification. 3) Inform all affected end users and/or sub-distributors and provide them with this notice. 4) Contact firm if you have affected product in inventory. - Discard any affected product(s) and confirm destruction. Or - Contact Hamilton Medical for a return material authorization number and return label. -Contact your sales representative to acquire replacement product. 5) Complete and return the Facility Reply Form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2463-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology