Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2465-2026

ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length. Recall

ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length. Catalog Number: iASB12-120. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

FDA recall Class II

Ongoing

FDA lists this Class II recall for ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2465-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 14, 2026
FDA report date
Jun 24, 2026
Quantity in commerce
891 packs (5,346 eaches)
Distribution
See key details
FDA cause
Process change control

What is being recalled?

ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length. Catalog Number: iASB12-120. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

Affected model & UDI checker

Search the exact identifiers published for Z-2465-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
iASB12-120ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length.Check source noticeNot specified
Not specifiedConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length.Check source notice10845854045725
FDA code information

Catalog Number: iASB12-120. UDI-DI: 10845854045725. Lot Numbers: 202507314, 202506194, 202510024, 202501165, 202504034, 202408225, 202411264. *OUS added 06/10/2026 expansion: 202510024, 202507314, 202504034, 202501165, 202411264, 202408225, 202506194. Manufactured from 14 June 2024 to 12 January 2026.

View the original FDA recall notice

Why was this product recalled?

Potential for overpressure alerts.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Conmed Corporation take?

ConMed notified consignees on about 05/14/2026 via emailed and mailed letter titled URGENT: Medical Device Correction. Consignees were made aware of the issue and risk associated with the issue. They were instructed that a replacement AirSeal disposable should be available prior to procedures in order to minimize potential delays. Before initiating insufflation, the user should ensure the tubing is properly connected between the AirSeal tube set and Airseal Access Ports. If the connection cannot be securely tightened and the overpressure alert cannot be resolved, please replace the affected device. Detailed instructions were provided on how to connect the AirSeal tube set and Airseal Access Ports. Consignees were also instructed to review inventory for affected units, place a copy of the provided Instructions for Use, Part No. AS-DIS-IFU, Rev. B with any inventory in possession with an active expiration date, contact all applicable departments and facilities that may have received affected products for awareness, and complete and return the medical device response form. If further information relevant to this issue becomes necessary CONMED will communicate as such. On about 06/04/2026 ConMed expanded their recall to include internationally distributed units. The letter sent included the same instructions.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology