STA - Fibrinogen 5. Recall
STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.
FDA recall Class II
Ongoing
FDA lists this Class II recall for STA - Fibrinogen 5.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2476-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 21, 2025
- FDA report date
- Sep 3, 2025
- Quantity in commerce
- 829,800 units (245,200 US)
- Distribution
- See key details
- FDA cause
- Mixed-up of materials/components
What is being recalled?
STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.
Affected model & UDI checker
Search the exact identifiers published for Z-2476-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 00674 | STA - Fibrinogen 5. | Check source notice | Not specified |
| Not specified | STA - Fibrinogen 5. | Check source notice | 13607450006749 |
FDA code information
Model/Catalog Number: 00674. UDI-DI: 13607450006749
Why was this product recalled?
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Diagnostica Stago, Inc. take?
Diagnostica Stago notified consignees on about 07/21/2025 via emailed letter titled URGENT: MEDICAL DEVICE RECALL (CORRECTION). Consignees were instructed to temporarily systematically run a quality control just before a batch of UFH or LMWH, if you run STA-LIQUID ANTI-Xa in UFH and LMWH applications on your STA SATELLITE analyzer. This request is being asked while Stago works to develop a permanent solution for the analyzer. Consignees were also requested to complete and return the Acknowledgement Form and to notify customers if affected units were transferred or further distributed.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2476-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology