Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2476-2025

STA - Fibrinogen 5. Recall

STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.

FDA recall Class II

Ongoing

FDA lists this Class II recall for STA - Fibrinogen 5.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2476-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 21, 2025
FDA report date
Sep 3, 2025
Quantity in commerce
829,800 units (245,200 US)
Distribution
See key details
FDA cause
Mixed-up of materials/components

What is being recalled?

STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.

Affected model & UDI checker

Search the exact identifiers published for Z-2476-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
00674STA - Fibrinogen 5.Check source noticeNot specified
Not specifiedSTA - Fibrinogen 5.Check source notice13607450006749
FDA code information

Model/Catalog Number: 00674. UDI-DI: 13607450006749

View the original FDA recall notice

Why was this product recalled?

Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Diagnostica Stago, Inc. take?

Diagnostica Stago notified consignees on about 07/21/2025 via emailed letter titled URGENT: MEDICAL DEVICE RECALL (CORRECTION). Consignees were instructed to temporarily systematically run a quality control just before a batch of UFH or LMWH, if you run STA-LIQUID ANTI-Xa in UFH and LMWH applications on your STA SATELLITE analyzer. This request is being asked while Stago works to develop a permanent solution for the analyzer. Consignees were also requested to complete and return the Acknowledgement Form and to notify customers if affected units were transferred or further distributed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.