DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113 Recall
DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113
FDA recall Class I
Ongoing
FDA lists this Class I recall for DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2480-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Aug 6, 2025
- FDA report date
- Sep 10, 2025
- Quantity in commerce
- 34549 units
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113
Affected model & UDI checker
Search the exact identifiers published for Z-2480-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 12113 | DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113 | Check source notice | Not specified |
| Not specified | DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113 | Check source notice | 20643169881338 |
| Not specified | DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113 | Check source notice | 00673978176475 |
| Not specified | DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113 | Check source notice | 00643169881334 |
FDA code information
UDI-DI: 20643169881338, 00673978176475, 00643169881334; Serial Numbers: 2023051188 2023060142 2023060144 2023060431 2023060762 2023061146 2023061146 2023070147 2023070147 2023070148 2023070148 2023070149 2023070149 2023070150 2023070150 2023070151 2023070490 2023070490 2023070979 2023080156 2023080797 2023080798 2023080798 2023081130 2023081131 2023081131 2023081132 2023081547 2023081548 2023081548 2023090234 2023090235 2023090427 2023090675 2023090979 2023090980 2023091104 2023091105 2023091105 2023091106 2023091106 2023100243 2023100243 2023100244 2023100641 2023100641 2023100642 2023101024 2023101025 2023101025 2023101365 2023101365 2023101366 2023110247 2023110308 2023110308 2023110312 2023110312 2023121249 2023121250 2023121250 2023121251 2023121251 2023121251 2023121252 2023121253 2023121254 2023121255 2024010470 2024010471 2024010472 2024010472 2024010473 2024010474 2024010474 2024010475 2024010475 2024010476 2024010476 2024010477 2024011016 2024011016 2024011214 2024031093 2024031094 2024031095 2024040067 2024040068 2024040068 2024040069 2024040070 2024040071 2024040245 2024040245 2024060793 202307C112 202308C248 202308C249 202309C022 202310C057 202310C058 202312C203 202402C086 202403C085 202403C086 202403C087 202406C057 202407C109
Why was this product recalled?
The catheters may not retain their shape.
What actions did MEDTRONIC PERFUSION SYSTEMS take?
Medtronic Cardiac Surgery issued an Urgent Medical Device Recall notice to its consignees on 08/06/2025 via UPS 2nd day delivery. The notice explained the issue with the device, potential harms if used, and requested the following: Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for listed product. " Immediately identify and quarantine all unused, listed product in your inventory. " " " Please share this notification with others in your organization as appropriate. If product listed above has been forwarded to another facility, please notify the facility of this . " Please maintain a copy of this communication in your records. Although the issue has been corrected for newly manufactured lots, please be aware that Medtronic will have limited product availability for these items over the next few months. If the product is unavailable, you may work with your sales representative to explore potential replacement options that Medtronic can offer. Alternatively, Medtronic will issue a credit note if a suitable replacement is not available. For questions, please contact your Medtronic Field Representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2480-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology