IntelliSpace Cardiovascular, Software 8.0.0.4. Recall
IntelliSpace Cardiovascular, Software 8.0.0.4.
FDA recall Class II
Ongoing
FDA lists this Class II recall for IntelliSpace Cardiovascular, Software 8.0.0.4.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2543-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 8, 2025
- FDA report date
- Sep 17, 2025
- Quantity in commerce
- 4 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
IntelliSpace Cardiovascular, Software 8.0.0.4.
Affected model & UDI checker
Search the exact identifiers published for Z-2543-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 830089 | IntelliSpace Cardiovascular, Software 8.0.0.4. | Check source notice | Not specified |
| Not specified | IntelliSpace Cardiovascular, Software 8.0.0.4. | Check source notice | 00884838122000 |
FDA code information
Model No. 830089; UDI: (01)00884838122000(11)240325(10)8.0.0.0; Serial No. 2210, 2050, 463, 193.
Why was this product recalled?
Software issue that results in the display of outdated information.
What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?
An URGENT Medical Device Correction notice dated 8/7/25 was mailed to consignees. The notification includes instructions for continued safe use of devices, which includes: 1. Ensure all finding codes with variables/modifiers are populated correctly on the report screen. Do not leave any FitB Modifier Code Fields Blank. 2. Verify report statements match the selections in the work area before finalizing report. Systems may continue to be used in accordance with the intended use and per the additional instructions. The provided recall notification is to be shared with all users of the device and retained with affected systems until the corrected software is installed. The provided recall response form should be returned within 30 days to Philips.recall@Philips.com. A Philips representative will call consignees to schedule a time to install the software solution (FCO83000218). Consignees with any questions can contact Philips Informatics Customer Support at 1-800-996-1328 (x1).
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2543-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology