Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Recall
Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Model/Catalog Number: 804-06-311 Product Description: Material: S.S/Nitronic, Non-Sterile Orthopedic instrument for shoulder prosthesis
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2546-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 18, 2025
- FDA report date
- Sep 17, 2025
- Quantity in commerce
- 927 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Model/Catalog Number: 804-06-311 Product Description: Material: S.S/Nitronic, Non-Sterile
Affected model & UDI checker
Search the exact identifiers published for Z-2546-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 804-06-311 | Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD | Check source notice | Not specified |
| Not specified | Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD | Check source notice | 00190446843825 |
FDA code information
Lot Code: Item Number: All Lots GTIN: 00190446843825
Why was this product recalled?
The reamer may kick or bind up during or immediately prior to use.
What actions did Encore Medical, LP take?
On July 29, 2025, Enovis issued a "Urgent Medical Device Recall Notification via Email to affected consignees. On August 4, 2025, Enovis issued an update to the recall notification via E-Mail requesting consignees rework affected trays and returned affected products. Enovis asked consignees to take the following actions: 1. Continue using the wedge reamer assembly while carefully inspecting and following the specific details in the supplied Quality Bulletin. 2. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 3. Please see the attached quality Bulletin 4. By 9/13/25, the firm will reach out regarding reworking and returning trays. 5. Acknowledge receipt of notification.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2546-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology