FDA RECALL RECORD · Z-2548-2026

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device Recall

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

FDA recall Class I

Ongoing

FDA lists this Class I recall for Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2548-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 21, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
10540 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

Affected model & UDI checker

Search the exact identifiers published for Z-2548-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedHillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing deviceCheck source notice00887761985018
FDA code information

UDI: 00887761985018, Lot Numbers: All lot numbers distributed beginning on 4/28/2025

View the original FDA recall notice

Why was this product recalled?

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

What actions did Baxter Healthcare Corporation take?

Baxter issued an URGENT MEDICAL DEVICE CORRECITON notice to its consignees on 5/21/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Before each use, carefully inspect the nebulizer cup for damage and ensure it is fully and securely locked during assembly, following the enclosed instructions. If leakage is observed during therapy, discontinue the use of that nebulizer and replace the patient circuit. 2. Nebulizer cups with patient circuits are routinely replaced as part of normal therapy usage. Additional patient circuits can be provided through standard distribution. 3. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any remaining inventory. Log in to the portal using the account number listed in the enclosed reply instruction sheet. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this urgent medical device correction in accordance with your customary procedures and check the associated box

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology