Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2555-2025

PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SP Recall

PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SP

FDA recall Class II

Ongoing

FDA lists this Class II recall for PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SP. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2555-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 24, 2025
FDA report date
Sep 17, 2025
Quantity in commerce
3
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SP

Affected model & UDI checker

Search the exact identifiers published for Z-2555-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
82-3713SPPILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SPCheck source noticeNot specified
Not specifiedPILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SPCheck source notice18257200161010
FDA code information

UDI-DI: 18257200161010, Lots: 001, 002, and 003

View the original FDA recall notice

Why was this product recalled?

A 37 mm implant width intervertebral body fusion device has a labeling discrepancy with incorrect width of 33 mm stated on the outer label side flap and inner label side flap information box. The 37 mm width on the front labels is correct.

What actions did Orthofix U.S. LLC take?

On 6/24/2025, recall notes were emailed to customers who were asked to do the following: 1) Examine your inventory and quarantine any affected product. 2) If implanted, healthcare Providers should determine whether it is necessary to inform the patient, based on clinical judgment and individual circumstances. 3) This notice should be shared with all relevant personnel within your organization and forwarded to any other organization or facility where potentially affected devices may have been transferred. 4) Complete and return the response form to the firm. Should you have any questions regarding this matter, please contact the firm at madison.ebel@seaspine.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.