Philips Telemetry Monitor 5500 1.4 GHz. Recall
Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips Telemetry Monitor 5500 1.4 GHz.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2579-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 5, 2026
- FDA report date
- Jul 1, 2026
- Quantity in commerce
- 909 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
Affected model & UDI checker
Search the exact identifiers published for Z-2579-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 867232 | Philips Telemetry Monitor 5500 1.4 GHz. | Check source notice | Not specified |
| Not specified | Philips Telemetry Monitor 5500 1.4 GHz. | Check source notice | 00884838102842 |
FDA code information
Model Number: 867232. UDI-DI: (01)00884838102842. Software Version Number: 4.0.0.0. Serial Numbers: US001A002F, US001A00AW, US001A00C9, US001A00EW, US001A00F0, US001A00GM, US001A00N4, US001A00PB, US001A00PZ, US001A00R1, US001A00UE, US001A00VU, US001A00Z4, US001A013T, US001A018A, US001A018R.
Why was this product recalled?
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Philips North America Llc take?
Philips Medical notified consignees on about 06/05/2026 via "URGENT Medical Device Correction" letter. Consignees were notified of the issue and potential hazard and instructed to continue the use of the device but avoid touching the display while the device is booting up and provide this notification to all personnel who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred. Philips will release new software which will correct this problem. A Philips representative will contact you to arrange for an upgrade to the new software.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2579-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology