Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2583-2025

i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300 Recall

i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300

FDA recall Class II

Ongoing

FDA lists this Class II recall for i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2583-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 21, 2025
FDA report date
Sep 24, 2025
Quantity in commerce
8,962,450 units
Distribution
See key details
FDA cause
No Marketing Application

What is being recalled?

i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

List Number: 03P88-25. All lot numbers.

View the original FDA recall notice

Why was this product recalled?

Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.

What actions did Abbott Point of Care, Inc. take?

Abbott notified consignees via email and letter on about 08/21/2025. Consignees were instructed to download, review and implement the updated i-STAT CG8+, EG7+ and EG6+ cartridge Instructions for Use (IFU), discontinue use of i-STAT CG8+, EG7+ and EG6+ cartridges for testing of capillary whole blood samples for potassium and ionized calcium, and use venous or arterial whole blood or an alternate method for testing capillary potassium and ionized calcium. Update the pH and PO2 reportable ranges on the i-STAT 1 analyzers running i-STAT CG8+, EG7+ or EG6+ cartridges. As this feature is not customizable through the i STAT 1 handheld keypad, please see the i STAT DE Quick Reference Guide - Customizing Reportable Ranges (ART 770547-00 Revision A) for more information. To use the Custom Reportable Range feature, you must have the i-STAT 1 handheld and i STAT DE version 2.3 or higher. Changes to the reportable range may require a change to the Laboratory Information System (LIS) interface. Additionally, consignees were requested to confirm receipt and understanding of this communication by responding to the business reply card included with the letter. If consignees have forwarded any i-STAT CG8+, EG7+ or EG6+ cartridges to another facility, they were requested to notify them.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology