Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder. Recall

Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2588-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 15, 2025
FDA report date
Sep 24, 2025
Quantity in commerce
12,471 units (6,565 US, 5,906 OUS)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.

Affected model & UDI checker

Search the exact identifiers published for Z-2588-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
HSK-3043Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.Check source noticeNot specified
Not specifiedHeartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.Check source notice00607567700321
FDA code information

Model No. HSK-3043; UDI: 00607567700321; Serial No. 3000379223, 3000382433, 3000385929, 3000391405, 3000399719, 3000404541, 3000406469, 3000408189, 3000411816, 3000414253, 3000414613, 3000415212, 3000419765, 3000421682, 3000422957, 3000427645, 3000428019, 3000429074, 3000438521, 3000442986, 3000444153, 3000444386, 3000451113, 3000451189, 3000454605, 3000457540, 3000459949, 3000459992, 3000465374, 3000466176, 3000466177, 3000471825, 3000473063, 3000473064, 3000474240, 3000474241, 3000475497, 3000478491, 3000478495, 3000482450, 3000485137, 3000488054 3000378620, 3000379223, 3000385929, 3000391405, 3000399719, 3000404541, 3000406469, 3000411816, 3000414253, 3000414613, 3000415212, 3000419765, 3000421682, 3000422957, 3000427645, 3000428019, 3000429074, 3000438076, 3000442986, 3000444153, 3000444386, 3000451113, 3000451189, 3000454605, 3000459949, 3000459992, 3000465374, 3000465479, 3000466177, 3000468296, 3000471825, 3000473064, 3000474240, 3000474241, 3000475497, 3000478491, 3000478495, 3000479466, 3000484006, 3000485137, 3000488054.

View the original FDA recall notice

Why was this product recalled?

Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.

What actions did Maquet Cardiovascular, LLC take?

Consignees were sent an URGENT MEDICAL DEVICE CORRECTION notification dated August 2025. The notification provides instructions for consignees to carry out in the instance of failure of the Heartstring Seal to load, failure of the Heartsring III Seal to deploy into the aortotomy, and failure of the deployed Heartstring III Seal to provide adequate hemostasis. Consignees are asked to forward the recall notification to individuals within their organization who use the recalled devices and to organizations to which product was further distributed. Consignees are to return the completed response form to recallresponses.qrc@getinge.com. Questions about this recall can be directed to Getinge Customer Service at 1-888-880-2874. ****Update 4/17/2026**** On August 15, 2025 URGENT MEDICAL DEVICE CORRECTION lupdate letters were sent to customers providing updated instructions for consignees to carry out in the instance of failure of the Heartstring Seal to load, failure of the Heartsring III Seal to deploy into the aortotomy, and failure of the deployed Heartstring III Seal to provide adequate hemostasis. Consignees are asked to forward the recall notification to individuals within their organization who use the recalled devices and to organizations to which product was further distributed. Consignees are to return the completed response form to recallresponses.qrc@getinge.com. Questions about this recall can be directed to Getinge Customer Service at 1-888-880-2874.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology