Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2595-2026

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. PICC LINEPACK (PCLUI)642-LF DYNJ47717A Recall

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. PICC LINEPACK (PCLUI)642-LF DYNJ47717A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. PICC LINEPACK (PCLUI)642-LF DYNJ47717A. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2595-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 27, 2026
FDA report date
Jul 1, 2026
Quantity in commerce
113, 843 kits
Distribution
See key details
FDA cause
No Marketing Application

What is being recalled?

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. PICC LINEPACK (PCLUI)642-LF DYNJ47717A

Affected model & UDI checker

Search the exact identifiers published for Z-2595-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. PICC LINEPACK (PCLUI)642-LF DYNJ47717ACheck source notice10193489454529
FDA code information

DYNJ47717A UDI-DI 10193489454529 Lots 21CMD153 21EMC033 21HMB521 21HMF278 21IME100 21IMG896 21JMG286 22AME235 22BMH056 22EMF125 22GMI215 22JME782 22LMC139 22LMC336

View the original FDA recall notice

Why was this product recalled?

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

What actions did Medline Industries, LP take?

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology