Philips CT systems labeled as the following: 1. Big Bore RT Recall
Philips CT systems labeled as the following: 1. Big Bore RT, Model Number: 728242. 2. Brilliance CT Big Bore Oncology, Model Number: 728243. 3. Brilliance CT Big Bore, Model Number: 728244.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips CT systems labeled as the following: 1. Big Bore RT. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2597-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 13, 2025
- FDA report date
- Oct 1, 2025
- Quantity in commerce
- 33 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Philips CT systems labeled as the following: 1. Big Bore RT, Model Number: 728242. 2. Brilliance CT Big Bore Oncology, Model Number: 728243. 3. Brilliance CT Big Bore, Model Number: 728244.
Affected model & UDI checker
Search the exact identifiers published for Z-2597-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 728242 | Philips CT systems labeled as the following: 1. Big Bore RT | Check source notice | Not specified |
| 728243 | Philips CT systems labeled as the following: 1. Big Bore RT | Check source notice | Not specified |
| 728244 | Philips CT systems labeled as the following: 1. Big Bore RT | Check source notice | Not specified |
| Not specified | Philips CT systems labeled as the following: 1. Big Bore RT | Check source notice | 00884838095168 |
| Not specified | Philips CT systems labeled as the following: 1. Big Bore RT | Check source notice | 00884838059450 |
FDA code information
1. Model Number: 728242; UDI-DI: 00884838095168; Serial Number: 766018. 2. Model Number: 728243; UDI-DI: N/A; Serial Numbers: 7179, 7238, 7273, 7418, 7467, 7608, 7634, 7618, 7801, 7925, 75111, 75045, 75119, 75035, 75132, 750019, 76023. 3. Model Number: 728244; UDI-DI: 00884838059450; Serial Numbers: 7271, 7979, 76234, 750003, 7878, 76054, 760019, 76135, 76230, 76147, 76171, 760057, 760058, 76270, 760075.
Why was this product recalled?
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
What actions did Philips North America Llc take?
Philips Medical notified consignees on about 08/11/2025 via letter. Consignees were instructed that they could continue using the affected systems accordance with the intended use and by following the recommendation: Observe the warning listed in Instructions for Use (IFU) and make sure that the patient is strapped securely to avoid dangling of the hands. Ensure that the patient is placed securely on the patient table and is not in danger of falling. Additionally, consignees were instructed to complete and return the provided response form, to circulate Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue and retain this letter with your system(s) until your system is inspected and corrected if necessary. Philips will contact consignees to schedule a time for a Philips Field Service Engineer (FSE) to visit each site and to inspect the affected system (FCO 72800832) and correct the system if necessary (FCO 72800833).
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2597-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026