FDA RECALL RECORD · Z-2602-2025

Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China Recall

Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model Number: 728324.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2602-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 13, 2025
FDA report date
Oct 1, 2025
Quantity in commerce
24 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model Number: 728324.

Affected model & UDI checker

Search the exact identifiers published for Z-2602-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
728324Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite ChinaCheck source noticeNot specified
Not specifiedPhilips CT systems labeled as the following: 1. Ingenuity Core128 / Elite ChinaCheck source notice00884838059863
FDA code information

1. Model Number: 728324; UDI-DI: 00884838059863; Serial Numbers: 336821, 337127, 336823, 336824, 336828, 336817, 336952, 336826, 336847, 336857, 336876, 336894, 336948, 336936, 336947, 337012, 337074, 337090, 337116, 337136, 345016, 345032, 345099, 345104.

View the original FDA recall notice

Why was this product recalled?

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

What actions did Philips North America Llc take?

Philips Medical notified consignees on about 08/11/2025 via letter. Consignees were instructed that they could continue using the affected systems accordance with the intended use and by following the recommendation: Observe the warning listed in Instructions for Use (IFU) and make sure that the patient is strapped securely to avoid dangling of the hands. Ensure that the patient is placed securely on the patient table and is not in danger of falling. Additionally, consignees were instructed to complete and return the provided response form, to circulate Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue and retain this letter with your system(s) until your system is inspected and corrected if necessary. Philips will contact consignees to schedule a time for a Philips Field Service Engineer (FSE) to visit each site and to inspect the affected system (FCO 72800832) and correct the system if necessary (FCO 72800833).

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology