Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2624-2026

Medtronic Harmony Delivery Catheter System Recall

Medtronic Harmony Delivery Catheter System

FDA recall Class I

Ongoing

FDA lists this Class I recall for Medtronic Harmony Delivery Catheter System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2624-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 22, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
1881 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Medtronic Harmony Delivery Catheter System

Affected model & UDI checker

Search the exact identifiers published for Z-2624-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMedtronic Harmony Delivery Catheter SystemCheck source notice00763000341367
Not specifiedMedtronic Harmony Delivery Catheter SystemCheck source notice00763000544027
Not specifiedMedtronic Harmony Delivery Catheter SystemCheck source notice00763000520151
Not specifiedMedtronic Harmony Delivery Catheter SystemCheck source notice00763000582951
Not specifiedMedtronic Harmony Delivery Catheter SystemCheck source notice00763000918712
FDA code information

GTIN 00763000341367, Lot Numbers: 0012508285, 0012517045, 0012526102, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663878, 0012663879, 0012690218, 0012699635, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824770, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965210, 0012965211, 0012976400, 0012986693, 0013063766, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624, 0013236643, 0013272796, 0013272798. GTIN 00763000544027, Lot Numbers: 0012508285, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663879, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965211, 0012976400, 0012986693, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624. GTIN 00763000520151, Lot Numbers: 0012575897, 0012642965, 0012728368, 0013272799. GTIN 00763000582951, Lot Numbers: 0012590535, 0012606411, 0012606413, 0012642964, 0012659418, 0012678282, 0012690219, 0012699637, 0012804151, 0012804580, 0012874527, 0012888682, 0012896885, 0012965221, 0012986699, 0013063764, 0013063765, 0013105037, 0013227504. GTIN 00763000918712, Lot Numbers: 0012635392, 0012676200, 0012690207, 0012874528, 0012874529, 0013015539, 0013105038.

View the original FDA recall notice

Why was this product recalled?

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medtronic Heart Valves Division take?

Medtronic issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 5/28/2026 via mail or email. The notice explained the issue, potential risk, and requested the following: " - Review your inventory for affected lot numbers listed. - If unused inventory from the listed lot numbers is located, immediately quarantine and return the affected units to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of the affected units. Your Medtronic sales representative can assist in the return of the affected units as necessary. - Please share this notification with implanters and teams within your organization. If product listed above has been forwarded from your facility to another facility, please notify the other facility of this Medtronic Urgent Medical Device Recall. - Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have the affected units in your possession. - Please maintain a copy of this communication in your records."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.